Voyance Auto Capture Troubleshooting

Rule Out Electromagnetic Interference (EMI)

Check for any source of EMI that could be triggering the Automatic Exposure Detection (AED) sensor:

  • Make sure the system is on a dedicated power outlet with proper grounding
  • Look for nearby high-powered equipment that may spike EMI

Step 2: Confirm Cable Integrity

  • Inspect all cables between the detector, generator, and acquisition PC
  • Check for kinks, frays, or bent connections
  • Make sure all connections are tight and clean

Step 3: Panel AED Sensitivity

  • Contact Maven Support